Common faults and troubleshooting methods for mini centrifuges

Mini centrifuge failure phenomenon, turn after the start, the head does not run:

Cause analysis and solution:

1. When the power cord is not plugged in, just plug in the power cord. 2. The power cord socket has no power supply. The solution ensures that the power supply works normally. 3. When the power switch is not turned on, the solution turns on the power switch. 4. When the centrifugal cover is not closed, just press down. Centrifugal cover 5. Please replace or inspect the supplier's technician when the fuse is burnt.
Mini Centrifuge failure phenomenon 2, after the start-up, the noise is large, severe vibration or abnormal noise:

Cause analysis and solution:

1. The instrument is not placed on a horizontal, sturdy platform. The solution is to place the instrument on a horizontal, sturdy platform. 2. The turret is not installed horizontally or in place, or the set screw is not locked. The solution is installed according to 4.b. 3. The test tube is not symmetrical. The distribution solution will place the test tube in equal symmetry. 4. The liquid level in the test tube is not equal. The solution is placed in the liquid level test tube. 5. The test tube is covered, and some samples are taken out, resulting in the rotor imbalance solution. Adding the sample to the balance 6. Test tube quality Poor, rupture leaks lead to rotor imbalance solution Replace with qualified test tube 7. Turn head aging or partial defect solution immediately stop using, replace the new rotor 8. Multi-standard test tube mixing solution Replace with the same specification test tube

Mini centrifuge failure phenomenon 3, the rotor speed is low or unstable:

Cause analysis and solution:

1. Excessive sample weight leads to overloading solution. The weight of each centrifuge tube must not exceed 3.0g.

2. Motor aging solution, please ask the supplier technician to repair

CRL-940.-,941

Fladrafinil CRL-40 941 Bisfluroadrafinil


CRL-40,941 is a molecule similar to adrafinil, and has been shown to be more effective than these two molecules in their indications, with fewer side effects.

Significant positive effects of CRL-40, 941 include improved memory, greater mental clarity, increased ambition, and overall higher levels of mental performance. As it improves focus, it could be used to treat conditions such as ADHD, Alzheimer's and depression. It also has great use in the treatment of sleep disorders, such as insomnia or narcolepsy.

Application Functions:

1. Improve memory

2. Increased learning ability

3. Improve cognitive processing

4. Improve your responsiveness

5. Improve your perception

6. Reduce anxiety

Reduce depression

What is the application of CRL-40 941?

* Applied in the pharmaceutical field.

* Used in the field of cosmetics.



Describe :

99% CAS 90212-80-9 Norootropic Fladrafinil CRL 40 941 CRL-40,941

Product Name

Fladrafinil CRL-40 941
Chemical Abstracts Society No.
90212-80-9
Appearance

White powder


Product photo :

Good CRL-40 Bisfluroadrafinil

Our company offers variety of products which can meet your multifarious demands.including API Powder.Pharmaceutical Intermediates.Vitamins Powder.Plant Extracts.Food Additive.Peptide Powder and so on We adhere to the management principles of "quality first, customer first and credit-based" since the establishment of the company and always do our best to satisfy potential needs of our customers. Our company is sincerely willing to cooperate with enterprises from all over the world in order to realize a win-win situation since the trend of economic globalization has developed with anirresistible force.

Crl-940.-,941,Fladrafinil Crl-40 941,Mt2 Nasal Spray,Stock 99% Noopept Nootropics

Shaanxi YXchuang Biotechnology Co., Ltd , https://www.peptide-nootropics.com